MDR technical documentation: what Annexes II and III require
What a medical device's technical documentation must contain under MDR Annexes II and III: description, design, GSPR, risk, V&V and post-market surveillance.
Key takeaways
- Technical documentation is the evidence file proving a medical device's conformity; its content is set by MDR Annexes II and III.
- Annex II has six parts: description and specifications, information supplied by the manufacturer, design and manufacturing, general safety and performance requirements, benefit-risk and risk management, verification and validation.
- Annex III covers post-market surveillance: the surveillance plan, the PSUR or the surveillance report.
- It is kept for at least 10 years after the last device is placed on the market, 15 years for implantables (Article 10(8)).
- The notified body assesses it for every class III and implantable class IIb device, and on a sampling basis for other class IIa and IIb devices.
Technical documentation is the file that demonstrates a medical device meets the MDR requirements. It is what the notified body assesses and what the competent authority may request at any time. MDR Annexes II and III set its content; this guide details them.
What technical documentation is for
MDR Article 10(4) requires the manufacturer to draw up and keep up to date technical documentation that allows the device's conformity to be assessed. It must be:
- complete: every required element is present or its absence is justified;
- structured and readily searchable: Annex II requires a clear, organised and unambiguous presentation;
- up to date: it follows changes to the device, standards and surveillance data;
- kept for at least 10 years after the last device covered by the EU declaration of conformity is placed on the market, 15 years for implantables (Article 10(8)).
Annex II: the six parts
1. Device description and specifications
- Trade name, general description, intended purpose, intended users
- Basic UDI-DI assigned by the manufacturer
- Patient population, indications, contraindications, warnings
- Principles of operation and mode of action
- Rationale for qualifying the product as a medical device
- Risk class and classification rule applied (Annex VIII)
- Novel features, accessories, variants and configurations
- Key components, materials in contact with the body, technical specifications
- References to previous generations and similar devices on the market
2. Information supplied by the manufacturer
Labels on the device and its packaging, and instructions for use, in the languages of the Member States where the device will be sold.
3. Design and manufacturing information
Design stages, manufacturing processes and their validation, continuous monitoring and final testing, and identification of all sites where design and manufacturing take place, including critical subcontractors and suppliers.
4. General safety and performance requirements
A requirement-by-requirement demonstration of conformity with Annex I:
- the applicable requirements, and justification for those that do not apply;
- the method used to demonstrate conformity;
- the harmonised standards, common specifications or other solutions applied;
- the precise reference to the documents providing the evidence.
This checklist is often the notified body's entry point into the assessment.
5. Benefit-risk analysis and risk management
The benefit-risk analysis and the solutions adopted, with the results of risk management under Annex I, usually based on ISO 14971.
6. Product verification and validation
Pre-clinical and clinical data:
- results of engineering, laboratory, simulated-use and, where applicable, animal tests;
- biocompatibility, electrical safety and electromagnetic compatibility;
- software verification and validation;
- stability and shelf life;
- clinical evaluation report and post-market clinical follow-up (PMCF) plan, or justification for its absence.
Additional information in specific cases: incorporated medicinal substance, tissues or cells of human or animal origin, substances absorbed by the body, CMR or endocrine-disrupting substances, sterile devices, measuring function, connection to other devices.
Annex III: post-market surveillance
The technical documentation also includes:
- the post-market surveillance plan (Article 84), with its data sources, indicators and thresholds, and the PMCF plan;
- the PSUR (classes IIa, IIb and III) or the post-market surveillance report (class I).
Our guide on post-market surveillance covers them in detail.
What the notified body assesses
| Class | Technical documentation assessment |
|---|---|
| IIa | On a representative sample, per category of devices |
| Non-implantable IIb | On a representative sample, per generic device group |
| Implantable IIb and III | For each device |
Language and access
The documentation is written in a language accepted by the notified body. The competent authority may request it, in full or in part, in an official EU language it determines. For a manufacturer established outside the EU, the authorised representative keeps a copy available to authorities.
Frequent findings
- A GSPR checklist without precise references to the evidence.
- An inconsistent intended purpose between the description, the instructions and the clinical evaluation.
- Standards cited in an outdated version, with no gap analysis against the current version.
- Manufacturing sites or critical subcontractors not identified.
- Documentation that does not follow device changes or PSUR conclusions.
Technical documentation ages with standards and MDCG guidance: keeping it current requires organised medical device regulatory monitoring.
Frequently asked questions
What does a medical device's technical documentation contain?
Under MDR Annex II: the device description and specifications, information supplied by the manufacturer (labelling, instructions for use), design and manufacturing information, demonstration of conformity with the general safety and performance requirements, benefit-risk analysis and risk management, and verification and validation, including clinical evaluation. Annex III adds post-market surveillance documentation.
How long must technical documentation be kept?
For at least 10 years after the last device covered by the EU declaration of conformity has been placed on the market, and at least 15 years for implantable devices (MDR Article 10(8)).
In which language must technical documentation be written?
In a language accepted by the notified body for assessment, and provided on request to the competent authority in an official EU language it determines. Labelling and instructions for use follow the languages required by each Member State where the device is made available.
What is the GSPR checklist?
It is the part of the documentation that lists each general safety and performance requirement of Annex I, states whether it applies (with a justification if not), the method used to demonstrate conformity, the harmonised standards or common specifications applied, and the reference to the evidence.
This topic keeps moving. So should your monitoring.
MedQara picks up every new official text (MDCG, EUR-Lex, ANSM, FDA, ISO) and flags what affects your devices. It is medical device regulatory monitoring, filtered on your scope and exportable for your ISO 13485 audits.
Create a free accountKeep receiving this kind of content
Once a week, receive analyses of new ANSM, FDA, EMA, MDR and IVDR publications - straight to your inbox.
No spam. Unsubscribe in 1 click. GDPR compliant.