CE marking of a medical device: the steps under the MDR
How to CE mark a medical device under the MDR: qualification, class, quality system, technical documentation, notified body, EU declaration, EUDAMED.
Key takeaways
- CE marking shows that a medical device meets the MDR; the manufacturer affixes it at the end of conformity assessment.
- The steps: qualify the product, classify it, choose the procedure, set up the quality system, build the technical documentation, appoint the PRRC and register in EUDAMED, go through notified body assessment, sign the EU declaration of conformity, affix the marking, register the device.
- Class I relies on self-declaration, except sterile, measuring or reusable devices; from class IIa a notified body is involved and its number accompanies the CE mark.
- A notified body certificate is valid for at most five years; the marking then commits the manufacturer to post-market surveillance and vigilance.
CE marking is a medical device's passport to the European Union market. It is not obtained from an administration: the manufacturer affixes it, after demonstrating conformity with Regulation (EU) 2017/745 through the procedure its class requires. Here are the steps, in the order they come.
1. Qualify the product
Is the product a medical device within the meaning of MDR Article 2(1)? It is if it has a medical purpose (diagnosis, prevention, monitoring, prediction, prognosis, treatment or alleviation of disease, injury or disability…) and its principal action is not achieved by pharmacological, immunological or metabolic means. For software, MDCG 2019-11 guides the reasoning; for borderline cases, the Commission's manual on borderline products.
2. Determine the class
The class (I, IIa, IIb or III) follows from the 22 rules of Annex VIII. It drives what comes next. Our guide on medical device classification presents them one by one.
3. Choose the conformity assessment procedure
Article 52 refers to Annexes IX to XI depending on the class:
- Class I: manufacturer's self-declaration, except sterile, measuring or reusable surgical devices, for which a notified body assesses those aspects;
- Class IIa: quality system (Annex IX) or product verification (Annex XI);
- Class IIb: Annex IX, or type examination (Annex X) with Annex XI;
- Class III: Annex IX, with technical documentation assessed for each device, or Annexes X and XI.
4. Set up the quality management system
Article 10(9) requires a quality system proportionate to the device class and type. The de facto reference is ISO 13485:2016, whose harmonised version gives presumption of conformity. It covers in particular the regulatory compliance strategy, risk management, clinical evaluation, post-market surveillance and change control.
5. Build the technical documentation
Annexes II and III: description, design and manufacturing, conformity with the Annex I general requirements, risk management under ISO 14971, verification and validation, clinical evaluation (Article 61, Annex XIV) and post-market surveillance plan. Our guide on technical documentation details the content.
6. Appoint the PRRC and register in EUDAMED
- Appoint a PRRC meeting the conditions of Article 15.
- Register the company in the EUDAMED Actor module to obtain its SRN, mandatory since 28 May 2026.
- Assign the device's Basic UDI-DI, before any notified body application.
- If the manufacturer is established outside the EU, designate an authorised representative (Article 11).
7. Go through notified body assessment
For the classes concerned, choose a notified body designated for your type of device (scope checkable in NANDO and EUDAMED), then apply. The assessment usually includes a quality system audit and a technical documentation assessment, on a sample basis or device by device depending on the class. For certain class III devices, an expert panel is consulted on the clinical evaluation (Article 54). At the end, the notified body issues a certificate, valid for at most five years (Article 56).
8. Draw up the EU declaration of conformity
The manufacturer drafts and signs the EU declaration of conformity (Article 19, Annex IV), taking responsibility for the device's conformity. It notably states the Basic UDI-DI, the class, the standards and common specifications applied and, where applicable, the notified body certificate.
9. Affix the CE marking
The CE marking is affixed visibly, legibly and indelibly to the device or its sterile packaging, to the instructions for use and to the sales packaging (Article 20, Annex V). Where a notified body was involved, its identification number appears next to it.
10. Register the device and prepare for life on the market
- Register the device in the EUDAMED UDI and device module.
- Prepare labelling and instructions in the languages required by each Member State of sale.
- Start post-market surveillance, PMCF and the vigilance system.
After marking: continuous conformity
CE marking commits the manufacturer to what follows: post-market surveillance, PSURs, vigilance, notified body surveillance audits, change control. Any significant change to the device must be assessed and, where required, submitted to the notified body before implementation.
Common mistakes
- Classifying too late: the class drives budget, timeline and choice of notified body.
- Choosing a notified body whose scope does not cover the device.
- Underestimating clinical evaluation, especially for class III and implantable devices.
- Forgetting EUDAMED: without an SRN, no notified body application.
- Treating marking as an end point: it opens an obligation of continuous surveillance.
Between the first step and the certificate, harmonised standards, MDCG guidance and transition rules may change: medical device regulatory monitoring means you do not discover them during an audit.
Frequently asked questions
How do you obtain CE marking for a medical device?
By demonstrating conformity with Regulation (EU) 2017/745 through a procedure that depends on the device class: quality system, technical documentation, clinical evaluation, then, except for class I devices that are non-sterile, non-measuring and not reusable, assessment by a notified body. The manufacturer then signs the EU declaration of conformity and affixes the marking.
How long does a medical device CE marking last?
The marking itself has no set duration, but the notified body certificate it relies on is valid for at most five years, renewable after a new assessment (MDR Article 56). The manufacturer must also maintain conformity continuously.
Who affixes the CE marking?
The manufacturer, under its sole responsibility, after conformity assessment. Where a notified body was involved, its identification number appears next to the CE marking.
Can a manufacturer outside the EU obtain CE marking?
Yes, provided it designates a single authorised representative established in the EU (MDR Article 11), who notably verifies the EU declaration of conformity and technical documentation and cooperates with authorities.
This topic keeps moving. So should your monitoring.
MedQara picks up every new official text (MDCG, EUR-Lex, ANSM, FDA, ISO) and flags what affects your devices. It is medical device regulatory monitoring, filtered on your scope and exportable for your ISO 13485 audits.